Philips Ingenia MRI System recalled due to incorrect pressure relief valve
Philips recalled 27 units of the Ingenia 1.5T Magnetic Resonance system with an incorrect pressure relief valve (42 psi instead of 3.75 psi) that could release helium gas and cause injury.
- Product
- Ingenia 1.5T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 7
- Category
- Distribution
- 12 states