The Recall Desk

FDA (Devices) recall action 88431

Siemens Healthcare Diagnostics, Inc recalled 3 products on 2021-09-15

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2021-09-15
Critical
0
Units
1,804

Why these products were recalled

Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot numbers

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2021-09-15

    Atellica IM Diagnostic Test Kit Recalled for Calibration Interval Deficiency

    Siemens Healthcare recalls the Atellica IM BR 27.29 diagnostic test kit because it does not maintain the claimed 10-day calibration interval; the required interval has been shortened to 3 days to ensure accurate results.

    Product
    Atellica IM BR 27.29 (BR) Assay 50 Test Kit- in vitro diagnostic use in the quantitative serial determination of cancer antigen CA 27.29 in human serum and plasma (EDTA) using the Atellica IM Analyzer. SMN: 10995478
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-09-15

    Medical Device Test Kit Requires More Frequent Calibration Than Labeled

    Siemens Atellica IM BR 27.29 test kit requires calibration every 3 days instead of the labeled 10-day interval. The product's calibration performance does not meet its Instructions for Use specifications.

    Product
    Atellica IM BR 27.29 (BR) Assay 250 Test Kit- in vitro diagnostic use in the quantitative serial determination of cancer antigen CA 27.29 in human serum and plasma (EDTA) using the Atellica IM Analyzer. SMN: 10995477
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-09-15

    Medical diagnostic kit requires more frequent calibration than Instructions stated

    The Instructions for Use for Siemens' Atellica IM BR 27.29 Test Kit incorrectly specified a 10-day calibration interval instead of 3 days. The manufacturer has updated the instructions for all kit lots.

    Product
    Atellica IM BR 27.29 (BR) Assay 250 Test Kit (Reference)- in vitro diagnostic use in the quantitative serial determination of cancer antigen CA 27.29 in human serum and plasma (EDTA) using the Atellica IM Analyzer.- SMN: 11206236
    Category
    Distribution
    0 states