The Recall Desk

FDA (Devices) recall action 88333

Philips Respironics, Inc. recalled 4 products on 2021-08-25

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2021-08-25
Critical
0
Units
10,000

Why these products were recalled

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2021-08-25

    Trilogy Evo Ventilator Software Defect May Cause Pressure Increase

    Philips Respironics recalls Trilogy Evo ventilators due to software issues that may increase pressure delivery to patients. The defect affects multiple product variants worldwide.

    Product
    Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X19, FR2110X14B, GB2110X15B, IA2110X15B, IN2110X15B, IN2110X19, IT2110X21B, KR2110X15B, LA2110X15B, LD2110X23B, ND2110X15B, RDE2110X13B, UDS2110X11B
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-08-25

    Trilogy Evo Respiratory Device Recalled for Software Pressure Issues

    Philips Respironics has recalled Trilogy Evo Universal respiratory devices (model DS2000X11B) for software defects that increase pressure, particularly in pediatric and infant applications.

    Product
    Trilogy Evo Universal, Product number DS2000X11B
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-08-25

    Trilogy Evo O2 Ventilator Software Issues Affecting Pressure Delivery

    Philips Respironics is recalling Trilogy Evo O2 ventilators with software issues that can increase pressure delivery. One issue affects infant/pediatric settings; another causes pressure drift during continuous use.

    Product
    Trilogy Evo O2, Product numbers DE2100X13B, DS2100X11B, EE2100X15B, ES2100X15B, EU2100X15B, FR2100X14B, FX2100X15B, IA2100X15B, IN2100X15B, IN2100X19, IT2100X21B, JP2100X16B, LA2100X15B, ND2100X15B, RDE2100X13B, SP2100X26B
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-08-25

    Trilogy EV300 Ventilator Recalled for Software-Related Pressure Calibration Issues

    Philips Respironics is recalling the Trilogy EV300 ventilator due to software issues causing pressure calibration errors affecting infant and pediatric patients. The defects impact 18,936 units distributed in the United States.

    Product
    Trilogy EV300, Product numbers CA2200X12B, DS2200X11B, FX2200X15B, IN2200X15B
    Category
    Distribution
    0 states