The Recall Desk

FDA (Devices) recall action 88219

BAXTER HEALTHCARE CORPORATION recalled 2 products on 2021-08-11

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-08-11
Critical
0
Units
5,102

Why these products were recalled

Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2021-08-11

    AMIA Automated Peritoneal Dialysis System User Guide Contains Incorrect Operating Instructions

    The AMIA Sharesource User Guide for the AMIA Automated Peritoneal Dialysis System incorrectly describes the Ultrafiltration Limit setting, potentially causing patients to misunderstand when therapy should end. Approximately 5,101 affected units have been distributed worldwide.

    Product
    Kaguya Automated Peritoneal Dialysis System
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2021-08-11

    Amia Automated PD System User Guide Incorrectly Describes Ultrafiltration Limit Function

    Baxter's user guide for the Amia Automated PD System incorrectly describes how the Ultrafiltration Limit setting controls the Extra Last Drain Mode. This documentation error could affect patients' ability to properly end therapy sessions.

    Product
    Sharesource Connectivity Platform for Use with the Amia Automated PD System
    Category
    Distribution
    0 states