The Recall Desk

FDA (Devices) recall action 88147

GE Healthcare, LLC recalled 14 products on 2021-07-21

The FDA (Devices) publishes a recall like this one as 14 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 14 pages.

Products
14
Announced
2021-07-21
Critical
0
Units
6,507

Why these products were recalled

GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–14 of 14

  • HighFDA (Devices)··2021-07-21

    MRI System Patient Orientation Button May Cause Flipped Images

    GE Healthcare's Discovery MR450 1.5T MRI system may inadvertently change patient orientation when users click the wrong button, resulting in flipped or incorrectly annotated images that could affect diagnostic accuracy.

    Product
    Discovery MR450 1.5T, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2021-07-21

    GE Healthcare MRI system button error causes incorrect diagnostic images

    GE Healthcare's SIGNA MR355 and MR360 MRI systems are recalled due to a button UI issue that may cause diagnostic images to be flipped or incorrectly annotated. No patient harm has been reported.

    Product
    SIGNA MR355, SIGNA MR360, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    MRI System User Interface Error May Cause Incorrectly Flipped Images

    GE Healthcare SIGNA Voyager MRI systems may display flipped or incorrectly annotated diagnostic images due to button confusion on the system interface. 541 devices are affected worldwide.

    Product
    SIGNA Voyager, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2021-07-21

    SIGNA MR380 MRI System Patient Orientation Button May Click Unintentionally

    GE Healthcare is recalling the SIGNA MR380 MRI system due to a user interface flaw where the Patient Orientation button may be inadvertently clicked instead of the Save RX button. This can result in incorrectly annotated or flipped MRI images.

    Product
    SIGNA MR380, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    GE SIGNA HDxt MRI System: Orientation Button Error Risk

    GE Healthcare recalls 91 SIGNA HDxt MRI systems worldwide. The Patient Orientation button may be accidentally clicked instead of Save RX, potentially causing diagnostic images to be incorrectly oriented or flipped.

    Product
    1.5T SIGNA HDxt (HD28) MR System, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    GE Healthcare SIGNA Pioneer MRI scanner control panel button design issue

    GE Healthcare recalls 460 SIGNA Pioneer MRI scanners worldwide due to a control panel button design flaw. Patient Orientation button may be accidentally clicked instead of Save RX, potentially resulting in incorrect image annotation or orientation.

    Product
    SIGNA Pioneer - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    MRI Scanner Patient Orientation Button Confusion Affects Images

    GE Healthcare's MRI scanners may have a button confusion issue where the Patient Orientation button is accidentally clicked instead of Save RX, potentially causing misoriented or incorrectly annotated images. 585 devices affected worldwide.

    Product
    Discovery MR750w 3.0T - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    GE Discovery MR750 MRI Scanner Button Error May Produce Flipped Images

    GE Healthcare recalled 248 units of the Discovery MR750 3.0T scanner due to a button-confusion error where the Patient Orientation button may be clicked instead of Save RX, potentially resulting in incorrectly oriented or flipped images. No injuries reported.

    Product
    Discovery MR750 3.0T - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    MRI Imaging System UI Defect May Cause Incorrect Image Orientation

    GE Healthcare recalled 1,737 SIGNA Creator and SIGNA Explorer MRI systems due to a software bug where the 'Patient Orientation' button can be accidentally clicked instead of 'Save RX,' potentially causing incorrectly oriented diagnostic images.

    Product
    SIGNA Creator, SIGNA Explorer, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    MRI System Patient Orientation Button Click Error May Flip Images

    GE Healthcare is recalling 1,519 Optima MR450w MRI systems worldwide due to a button interface error. The 'Patient Orientation' button may be unintentionally clicked instead of 'Save RX,' causing incorrectly oriented or flipped diagnostic images.

    Product
    Optima MR450w 1.5T, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    SIGNA Artist MRI System Patient Orientation Button May Cause Flipped Images

    GE Healthcare is recalling 562 SIGNA Artist MRI systems due to a user interface defect where the Patient Orientation button may be inadvertently clicked instead of the Save RX button, potentially causing images to be incorrectly annotated or flipped.

    Product
    SIGNA Artist, Nuclear Magnetic Resonance Imaging System
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    MRI Scanner Patient Orientation Button Confusion May Flip Diagnostic Images

    GE Healthcare recalls 311 SIGNA Architect MRI scanners due to button confusion that may result in flipped diagnostic images. The 'Patient Orientation' button may be accidentally clicked instead of 'Save RX', causing incorrect image orientation.

    Product
    SIGNA Architect - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    GE Healthcare SIGNA PET/MR button mix-up causes image orientation errors

    GE Healthcare is recalling 71 SIGNA PET/MR imaging devices worldwide. A software issue allows the Patient Orientation button to be accidentally clicked instead of Save RX, resulting in incorrectly annotated or flipped diagnostic images.

    Product
    SIGNA PET/MR, Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-21

    MRI scanner user interface defect may cause incorrectly oriented images

    GE Healthcare recalls SIGNA Premier MRI scanners (100 devices, worldwide distribution) due to a user interface defect where the Patient Orientation button may be inadvertently clicked instead of Save RX, potentially resulting in incorrectly annotated or flipped diagnostic images.

    Product
    SIGNA Premier - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
    Category
    Distribution
    0 states