The Recall Desk

FDA (Devices) recall action 88087

Boston Scientific Corporation recalled 6 products on 2021-07-14

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2021-07-14
Critical
0
Units
138,627

Why these products were recalled

Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–6 of 6

  • HighFDA (Devices)··2021-07-14

    VISIONIST CRT-P Pacemaker Recalled Due to Battery Depletion Risk

    Boston Scientific has recalled 5,920 VISIONIST CRT-P pacemakers due to hydrogen gas buildup that can compromise a capacitor and cause accelerated battery depletion. Affected patients should contact their healthcare provider for device evaluation.

    Product
    VISIONIST CRT-P pacemaker, Models: a) U225 b) U226 c) U228
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-14

    Proponent pacemakers recalled for accelerated battery depletion risk

    Boston Scientific has recalled certain Proponent pacemaker models worldwide. Hydrogen gas may accumulate inside the device, compromising the capacitor and causing accelerated battery depletion.

    Product
    Proponent pacemaker, Models: a) L200 b) L201 c) L209 d) L210 e) L211 f) L221 g) L231
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-14

    VALITUDE CRT-P Pacemaker Recalled Due to Potential Accelerated Battery Depletion

    Boston Scientific VALITUDE CRT-P pacemakers (models U125 and U128) may experience hydrogen release that compromises the capacitor, causing accelerated battery depletion. Patients should consult their healthcare provider.

    Product
    VALITUDE CRT-P pacemaker, Models: a) U125 b) U128
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-14

    Altrua 2 Pacemakers at Risk from Hydrogen-Induced Battery Failure

    Boston Scientific is recalling 2,952 Altrua 2 pacemaker units (models S701, S702, S722) distributed worldwide due to hydrogen release that can damage the capacitor and accelerate battery depletion.

    Product
    Altrua 2 pacemaker, Models: a) S701 b) S702 c) S722
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-14

    Essentio Pacemaker Models Recalled for Potential Battery Depletion Risk

    Boston Scientific is recalling six Essentio pacemaker models due to a hydrogen release defect that may compromise a critical capacitor and cause accelerated battery depletion. Approximately 31,548 units worldwide are affected.

    Product
    Essentio pacemaker, Models: a) L100 b) L101 c) L110 d) L111 e) L121 f) L131
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-07-14

    Accolade Pacemakers Recalled for Hydrogen-Related Battery Depletion

    Boston Scientific is recalling Accolade pacemaker models L300, L301, L310, L311, L321, and L331 due to latent hydrogen release that compromises the capacitor and accelerates battery depletion.

    Product
    Accolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331
    Category
    Distribution
    0 states