FDA (Devices) recall action 87960

Beckman Coulter Inc. recalled 2 products on 2021-06-30

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-06-30
Critical
0
Units
17,180

Why these products were recalled

SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2021-06-30

    SARS-CoV-2 IgG II Calibrator Results Potentially Multiplied by 1000

    Beckman Coulter's Access SARS-CoV-2 IgG II Calibrator may return test results inflated by up to 1000-fold due to software errors, potentially misrepresenting patient immune status.

    Product
    Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751
    Category
    Distribution
    31 states