FDA (Devices) recall action 87882

Nico Corp. recalled 2 products on 2021-06-16

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-06-16
Critical
0
Units
519

Why these products were recalled

(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the pati…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2021-06-16

    Myriad-LX Light Source Recalled Due to Burn and Occlusion Risk

    Nico Corp. is recalling 80 Myriad-LX Light Source units due to potential burn injury to patients from the distal tip of the fiber assembly and risk of delivery sleeve occlusion during surgical use.

    Product
    Myriad-LX Light Source-for delivery of light to the surgical field to enhance visualization of tissue during resection Part Number NN-7013 The Myriad-LX light source is a high-intensity 300W Xenon illuminator with a touch screen and digital dimming controls. The source is…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-06-16

    Medical Device Illumination Pack Recalled Due to Burn and Occlusion Risks

    Nico Corp. is recalling 439 NICO Myriad-LX Illumination Packs used in surgical procedures due to potential delivery sleeve occlusion and thermal burn injury from the fiber tip assembly.