FDA (Devices) recall action 87705

Roche Diagnostics Operations, Inc. recalled 2 products on 2021-05-26

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-05-26
Critical
0
Units
726

Why these products were recalled

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2021-05-26

    Roche Recalls Troponin I Immunoassay Kits Due to False Negative Risk

    Roche Diagnostics is recalling 726 Elecsys Troponin I Immunoassay kits because they may produce false negative results when run immediately after proBNP II tests on specific modules.

    Product
    Elecsys Troponin I Immunoassay: Catalog# 05094810160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
    Category
    Distribution
    28 states
  • HighFDA (Devices)··2021-05-26

    Roche Recalls Troponin I STAT Immunoassay Kits Due to False Negative Risk

    Roche Diagnostics is recalling 10,489 Troponin I STAT Immunoassay kits because they may produce false negative results when run after proBNP II tests on specific modules.

    Product
    Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
    Category
    Distribution
    28 states