The Recall Desk

FDA (Devices) recall action 87654

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalled 2 products on 2021-05-12

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-05-12
Critical
0
Units
61,722

Why these products were recalled

There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2021-05-12

    Medtronic Recalls CareLink SmartSync Device Manager Due to Software Error

    Medtronic is recalling 16,311 CareLink SmartSync Device Managers because a software issue may overestimate the remaining battery life of implanted cardiac devices.

    Product
    CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-05-12

    Medtronic Recalls CRT-P Quad Models Due to Software Longevity Error

    Medtronic is recalling 45,411 CRT-P Quad implantable devices worldwide because a software issue overestimates battery longevity for six months before replacement is due.

    Product
    CRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implantable medical devices used to treat arrhythmias related to bradycardia, and…
    Category
    Distribution
    0 states