The Recall Desk

FDA (Devices) recall action 87446

Heartware, Inc. recalled 6 products on 2021-04-14

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2021-04-14
Critical
0
Units

Why these products were recalled

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

The agency’s own wording, quoted from the enforcement report.

1–6 of 6

  • ModerateFDA (Devices)··2021-04-14

    Medtronic HVAD Battery Recall for Potential Controller Port Damage

    Heartware, Inc. is recalling Medtronic HVAD Batteries because controller power and data cables may damage controller ports due to wear from misalignment.

    Product
    Medtronic HVAD Battery, Product Number: 1650, 1650CA-CLIN, 1650DE
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2021-04-14

    Medtronic HVAD Alarm Adapter Recalled for Potential Port Damage

    Heartware, Inc. is recalling 418 Medtronic HVAD Alarm Adapters because cable misalignment may cause damage to the HVAD Controller ports.

    Product
    Medtronic HVAD Alarm Adapter, Product Number: 1450
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2021-04-14

    Medtronic HVAD Monitor Recall Due to Cable Port Damage Risk

    Heartware is recalling Medtronic HVAD Packaged Monitors because worn cables may damage controller ports during insertion.

    Product
    Medtronic HVAD Packaged Monitor, Model Numbers: 1500, 1500AU, 1500DE, 1510US, 1511AR, 1511AU, 1511BR, 1511CA, 1511CH, 1511DE, 1511GB, 1511IL, 1511IN, 1511IT, 1511JP, 1511MY, 1520CLIN-AU, 1520CLIN-DE, 1520CLIN-GB, 1520JP, 1520US, 1521AU, 1521BR, 1521CA, 1521CH, 1521DE, 1521GB, 15
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-04-14

    Medtronic HVAD Controller Power Cables Recalled for Port Damage Risk

    Heartware, Inc. is recalling Medtronic HVAD Controller power and data cables due to a risk of damaging controller ports from cable misalignment.

    Product
    Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 14
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2021-04-14

    Medtronic HVAD Controller Recall for Potential Port Damage

    Heartware is recalling Medtronic HVAD Controllers because power and data cables may damage controller ports due to wear from misalignment.

    Product
    Medtronic HVAD Controller, Product Numbers: 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB, 1401IL, 1401IN, 1401IT, 1401JP, 1401US, 1403US, 1407AR, 1407AU, 1407BR, 1407CA, 1407CA-CLIN, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR
    Category
    Distribution
    0 states