The Recall Desk

FDA (Devices) recall action 86803

B Braun Medical, Inc. recalled 3 products on 2020-12-30

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2020-12-30
Critical
0
Units
349

Why these products were recalled

There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2020-12-30

    B Braun APEX Compounding System Control Panel Module Recalled for Accuracy Issues

    B Braun Medical is recalling 248 APEX Compounding System Control Panel Modules because they may fail to detect air bubbles, potentially resulting in inaccurate drug compounding.

    Product
    APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.
    Category
    Distribution
    27 states
  • HighFDA (Devices)··2020-12-30

    B Braun APEX Compounding System Control Panel Module Recalled

    B Braun Medical is recalling 16 APEX Compounding System Control Panel Modules because they may fail to detect air bubbles, potentially compromising drug accuracy.

    Product
    APEX Compounding System Control Panel Module, REF: 601238, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.
    Category
    Distribution
    27 states
  • HighFDA (Devices)··2020-12-30

    B Braun APEX Compounding System Control Panel Module Recalled

    B Braun Medical is recalling 85 APEX Compounding System Control Panel Modules because they may fail to detect air bubbles, potentially resulting in inaccurate drug compounding.

    Product
    APEX Compounding System Control Panel Module, REF: 601237, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.
    Category
    Distribution
    27 states