Siemens Healthcare Diagnostics, Inc. recalled 2 products on 2020-11-25

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2020-11-25
Critical
0
Units
—

Why these products were recalled

Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values

Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values

The agency’s own wording, quoted from the enforcement report.

1–2 of 2

  • ModerateFDA (Devices)··2020-11-25

    Siemens Atellica IM 1600 Analyzer Recalled for Custom Settings Reset

    Siemens Healthcare Diagnostics is recalling Atellica IM 1600 Analyzers because scanning a new kit lot may reset custom settings to defaults.

    Product
    Atellica IM 1600 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066000
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-25

    Siemens Atellica IM 1300 Analyzer Recalled for Test Definition Scanning Issues

    Siemens Healthcare Diagnostics is recalling Atellica IM 1300 Analyzers because scanning specific test definitions may reset custom settings, causing inaccurate test results.

    Product
    Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001