The Recall Desk

FDA (Devices) recall action 86581

Inpeco S.A. recalled 3 products on 2020-11-18

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2020-11-18
Critical
0
Units
333

Why these products were recalled

When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this …

When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2020-11-18

    Inpeco Accelerator a3600 Aliquoter Module Recall for Sample Dilution Risk

    Inpeco S.A. is recalling 136 Accelerator a3600 Automation Systems with Aliquoter Modules due to a software error that may dilute samples with water, leading to erroneous test results.

    Product
    Accelerator a3600 Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample proc
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-18

    Inpeco FlexLab Automation System Recalled for Potential Sample Dilution

    Inpeco S.A. is recalling 93 FlexLab Automation Systems with Aliquoter Modules because a software error may cause samples to be diluted with water, leading to erroneous test results.

    Product
    FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processin
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-18

    Inpeco Aptio Automation System Recall for Potential Sample Dilution

    Inpeco S.A. is recalling 104 Aptio Automation Systems with Aliquoter Modules because a software error may cause samples to be diluted with water, leading to erroneous test results.

    Product
    Aptio Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in th
    Category
    Distribution
    0 states