ICU Medical, Inc. recalled 13 products on 2020-12-30

The FDA (Devices) publishes a recall like this one as 13 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 13 pages.

Products
13
Announced
2020-12-30
Critical
0
Units
3,083,479

Why these products were recalled

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–13 of 13

  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Spiros Male Luer Connectors Due to Molding Defect

    ICU Medical is recalling Spiros male luer connectors and administration sets due to a molding defect that may cause leaks, potentially leading to fluid path contamination or exposure to hazardous medications.

    Product
    30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Spiros, Closed Male Luer, REF: CH2000S; Spinning Spiros Closed Male Luer, 10 units, REF: CH2000S-10…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Oncology and Administration Sets Due to Molding Defect

    ICU Medical is recalling specific oncology and administration sets because a molding defect in male luers may cause leaks, potentially leading to fluid path contamination or exposure to hazardous medications.

    Product
    76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Spiros Blood Sets Due to Molding Defect

    ICU Medical is recalling Spiros blood sets because a molding defect may cause leaks, risking fluid contamination or air-in-line.

    Product
    26 IN(66 cm) APPX 3.4ml, 10 DROP BLOOD SET, 170 MICRON FILTER, Spiros, REF: Z2795
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls ChemoLock Admin Sets Due to Molding Defect

    ICU Medical is recalling specific ChemoLock administration sets because a molding defect may cause leaks. This could lead to fluid path contamination or exposure to hazardous medications.

    Product
    30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL3011T; 40" (102 cm) Appx 4.9 ml, Admin Set…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Spiros Administration Sets Recalled for Molding Defect

    ICU Medical is recalling Spiros administration sets due to a molding defect that may cause leaks, potentially leading to fluid path contamination or exposure to hazardous medications.

    Product
    42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, REF: CH3330-C; 60 IN (152 cm) Appx 1.1 ml, Smallbore Set, MicroClave Clear, Spiros, REF: CH3353; 8"…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls 30-inch IV Administration Sets Due to Molding Defect

    ICU Medical is recalling 325 units of 30-inch IV administration sets because a molding defect in the male luer may cause fluid leaks, potentially leading to infusion delays or contamination.

    Product
    30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber, 1.2 Micron Filter, Spiros, Bag Hanger, REF: CH3286
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Appx 1.7 ml Bag Spike Due to Molding Defect

    ICU Medical is recalling 900 units of Appx 1.7 ml Bag Spike w/Integrated Clave due to a molding defect that may cause leaks in male luers.

    Product
    Appx 1.7 ml Bag Spike w/Integrated Clave, 2 Drop-In Spiros, REF: CH3260
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Spinning Spiros and ChemoAccess Infusion Components

    ICU Medical is recalling specific infusion components due to a molding defect that may cause leaks, risking fluid contamination or exposure to hazardous medications.

    Product
    Spinning Spiros, Closed Male Luer, REF: 060-CH2000S; SURPLUG ChemoAccess Closed Male Luer, REF: 066-CH2000S; SURPLUG ChemoAccess 5" (13cm) Bag Spike Adaptor with Closed Male Luer, REF: 066-CH3034; 35" (89 cm) Appx 3.9 ml, 10 Drop Admin Set w/Integrated Clave Drip Chamber, Clave…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Spinning Spiros Closed Male Luer Components

    ICU Medical is recalling specific Spinning Spiros components due to a molding defect that may cause leaks, risking fluid path contamination or exposure to hazardous medications.

    Product
    Spinning Spiros Closed Male Luer, Red Cap, REF: IB-CH2000SC; 5" (13 cm) Appx 1.5 ml, Add-On Set w/Spiros, REF: IB-CH3034; 5" (13 cm) Appx 1.5 ml, Bag Spike Adapter w/Spiros, Vented Cap, REF: IB-CH3034C
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Spinning Spiros Syringes and Administration Sets

    ICU Medical is recalling specific syringes and administration sets with spinning Spiros components due to a molding defect that may cause fluid leaks.

    Product
    3mL Syringe w/Spinning Spiros, Red Cap, REF: CH2003; 5mL Syringe w/Spinning Spiros, Red Cap, REF: CH2005; 10mL Syringe w/Spinning Spiros, Red Cap, REF: CH2010; 20mL Syringe w/Spinning Spiros, Red Cap, REF: CH2020; 30mL Syringe w/Spinning Spiros, Red Cap, REF: CH2030; 60mL…
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-12-30

    ICU Medical Recalls ChemoLock Administration Sets Due to Molding Defect

    ICU Medical is recalling specific ChemoLock administration sets because a molding defect may cause leaks. This could lead to fluid path contamination or exposure to hazardous medications.

    Product
    31" (78 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3011; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3020; ChemoLock Port w/Spiros, Red Cap, REF: CL-35…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Spiros Male Luer Devices Due to Molding Defect

    ICU Medical is recalling Spiros closed male luer devices due to a molding defect that may cause leaks. The defect could lead to infusion delays, fluid contamination, or exposure to hazardous medications.

    Product
    Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-12-30

    ICU Medical Recalls Spiros and Clave IV Administration Sets

    ICU Medical is recalling specific Spiros and Clave IV administration sets due to a molding defect that may cause leaks in spinning male luer components.

    Product
    5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135…
    Category
    Distribution