The Recall Desk

FDA (Devices) recall action 86263

Inpeco S.A. recalled 3 products on 2020-09-23

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2020-09-23
Critical
0
Units

Why these products were recalled

Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2020-09-23

    Inpeco FlexLab Laboratory Automation System Recalled for Sample ID Errors

    Inpeco S.A. is recalling FlexLab laboratory automation systems because firmware issues may cause sample ID mis-association, leading to incorrect or delayed test results.

    Product
    FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Laboratory Automation system - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Pos
    Category
    Distribution
    12 states
  • SevereFDA (Devices)··2020-09-23

    Inpeco Aptio Automation Systems Recalled for Potential Sample ID Mis-association

    Inpeco S.A. is recalling Aptio Automation systems with Sysmex Interface Modules because a firmware defect may mis-associate sample IDs, leading to incorrect or delayed clinical chemistry results.

    Product
    Aptio Automation systems with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analyti
    Category
    Distribution
    12 states
  • HighFDA (Devices)··2020-09-23

    Inpeco FlexLab 3.6 Automation Recalled for Potential Sample ID Mis-association

    Inpeco S.A. is recalling 8 FlexLab 3.6 Automation systems because firmware issues may cause sample ID mis-association, leading to incorrect or delayed clinical chemistry results.

    Product
    Accelerator a3600 system with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analyti
    Category
    Distribution
    12 states