The Recall Desk

FDA (Devices) recall action 85737

Siemens Healthcare Diagnostics, Inc. recalled 2 products on 2020-07-08

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2020-07-08
Critical
0
Units
256

Why these products were recalled

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2020-07-08

    Siemens CentraLink Software Recalled for Potential Erroneous Patient Results

    Siemens Healthcare Diagnostics is recalling CentraLink Data Management System software versions 16.0.2 and 16.0.3 due to a defect that may cause erroneous patient results.

    Product
    (1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-07-08

    Siemens Atellica Data Manager Software Recalled for Potential Erroneous Patient Results

    Siemens Healthcare Diagnostics is recalling Atellica Data Manager software versions 1.0 and 1.1 because unexpected interface driver behavior may cause quality control results to be assigned to incorrect control lot numbers, potentially leading to the reporting of erroneous patient results.