FDA (Devices) recall action 85634

Radiometer Medical ApS recalled 2 products on 2020-07-08

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2020-07-08
Critical
0
Units
11,354

Why these products were recalled

Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2020-07-08

    Radiometer ABL80-FLEX CO-OX Box Label Recall for Barcode Misinterpretation

    Radiometer Medical ApS is recalling ABL80-FLEX CO-OX box labels because barcode misinterpretation on analyzers may cause patient mix-ups or delayed treatment.

    Product
    ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
    Category
    Distribution
    49 states
  • HighFDA (Devices)··2020-07-08

    Radiometer ABL80-FLEX Box Label Recall for Barcode Misinterpretation Risk

    Radiometer Medical ApS is recalling ABL80-FLEX box labels because barcode misinterpretation on analyzers may cause patient mix-ups or delayed treatment.

    Product
    ABL80-FLEX-393-839 Box Label, REF 393-839, IVD, CE, UDI: 05700693938394 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
    Category
    Distribution
    49 states