The Recall Desk

FDA (Devices) recall action 85241

Philips North America, LLC recalled 2 products on 2021-02-03

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-02-03
Critical
0
Units
103,298

Why these products were recalled

The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2021-02-03

    Philips Telemetry Transceiver Recall for Heart Rate Display Failure

    Philips is recalling TRx4841A telemetry transceivers because they may fail to display heart rates or alarms when used with specific Patient Information Center software.

    Product
    TRx4841A 1.4 GHz IntelliVue Tele TRX, Model Number 862439; Includes System Codes 453564007261, 453564007271, 453564007281, 989803196951
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-02-03

    Philips Telemetry Transceiver Recall for Heart Rate Display Failure

    Philips is recalling 21,672 telemetry transceivers because they may fail to display heart rates or generate alarms when used with specific Patient Information Center software.