Philips Medical Systems Gmbh, DMC recalled 2 products on 2020-02-26

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2020-02-26
Critical
0
Units
247

Why these products were recalled

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2020-02-26

    Philips CombiDiagnost GCF X-ray System Software Recall

    Philips is recalling 116 CombiDiagnost GCF X-ray systems due to a software error that locks the table and an incorrectly installed thermos switch.

    Product
    CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
    Category
    Distribution
    18 states
  • HighFDA (Devices)··2020-02-26

    Philips CombiDiagnost PCF X-ray System Software Recall

    Philips is recalling 131 CombiDiagnost PCF X-ray systems due to a software error that locks the table and an incorrectly installed thermos switch.

    Product
    CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
    Category
    Distribution
    18 states