The Recall Desk

FDA (Devices) recall action 84741

Delfi Medical Innovations, Inc. recalled 2 products on 2020-05-13

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2020-05-13
Critical
0
Units
83

Why these products were recalled

Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh di…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2020-05-13

    Delfi VariFit Contour Cuffs Recalled for Inadequate Weld

    Delfi Medical Innovations is recalling 59 units of VariFit Contour Cuffs due to a manufacturing defect that may cause the cuff to leak.

    Product
    Delfi Medical Innovations, VariFit Contour Cuffs, single use adult thigh cuff, single port, single bladder, REF 9-9300-003. Rx Only, Sterile
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2020-05-13

    Delfi VariFit Sterile Thigh Tourniquet Cuffs Recalled for Weld Defect

    Delfi Medical Innovations is recalling certain lots of VariFit Sterile Single Use Thigh Cuffs due to a potential weld defect that may cause the cuff to leak.