The Recall Desk

FDA (Devices) recall action 84544

Palodex Group Oy recalled 4 products on 2020-01-29

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2020-01-29
Critical
0
Units
9,524

Why these products were recalled

Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • SevereFDA (Devices)··2020-01-29

    KaVo Scan eXam X-Ray Imager Recalled for Power Supply Fire Risk

    Palodex Group Oy is recalling KaVo Scan eXam X-ray imagers because plastic insulation degradation in the power supply may cause short circuits, heating, and smoke.

    Product
    KaVo Scan eXam SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2020-01-29

    Palodex Dental X-Ray Imagers Recalled for Power Supply Fire Risk

    Palodex Group Oy is recalling Instrumentarium Dental Express X-ray imagers because slow degradation of plastic insulation in the power supply unit may cause short circuits, heating, and smoke.

    Product
    Instrumentarium Dental Express SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2020-01-29

    Soredex DIGORA Optime X-Ray Imager PSU Insulation Recall

    Palodex Group Oy is recalling 297 Soredex DIGORA Optime X-ray imagers because plastic insulation in the power supply may degrade, causing short circuits, heating, and smoke.

    Product
    Soredex DIGORA Optime SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2020-01-29

    Gendex X-Ray Imager Recall Due to Power Supply Insulation Degradation

    Palodex Group Oy is recalling Gendex GXPS-500 X-ray imagers because plastic insulation in the power supply may degrade, causing short circuits, heating, and smoke.

    Product
    Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
    Category
    Distribution
    2 states