FDA (Devices) recall action 84450

Nobel Biocare Usa Llc recalled 24 products on 2020-02-26

The FDA (Devices) publishes a recall like this one as 24 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 24 pages.

Products
24
Announced
2020-02-26
Critical
0
Units
5,689

Why these products were recalled

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–24 of 24

  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Abutment Screw Retrieval Kits Due to Sterility Concerns

    Nobel Biocare is recalling 7 units of Abutment Screw Retrieval Kits because a packaging sealing issue may compromise the sterility of the included medical devices.

    Product
    Abutment Screw Retrieval Kit - Product Usage: The kit itself does not have a specific intended Use, but included devices have specific intended use as shown above.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Guided Start Drills Due to Sterility Concerns

    Nobel Biocare is recalling Guided Start Drills because a packaging sealing issue may compromise the sterility of the devices during their shelf life.

    Product
    Guided Start Drills/Counterbore NobRpl NP - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Dental Drills Recalled Due to Sterility Concerns

    Nobel Biocare is recalling specific dental drills because a packaging defect may compromise their sterility during shelf life.

    Product
    Drill Tapered 6.0 6x8 mm Single-Pat - Product Usage: Drills are intended to be used in the upper and lower jaw bone to prepare osteotomy prior to implant placement.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Dense Bone Drill Recall Due to Sterility Concerns

    Nobel Biocare is recalling Dense Bone Drills because a packaging sealing issue may compromise sterility during shelf life.

    Product
    Dense Bone Drill NP 3.5x16 mm Single-Pat - Product Usage: Drills are intended to be used in the upper and lower jaw bone to prepare osteotomy prior to implant placement.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Tapered Drills Recalled Due to Sterility Concerns

    Nobel Biocare is recalling specific tapered dental drills because a packaging defect may compromise their sterility during shelf life.

    Product
    Drill Tapered NP 3.5x10 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Guided Tapered Drills Recalled for Sterility Concerns

    Nobel Biocare is recalling 346 Guided Tapered Drills because a packaging sealing issue may compromise sterility during shelf life.

    Product
    Drill Tapered RP 4.3x10 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Screw Taps Due to Packaging Seal Defects

    Nobel Biocare is recalling Screw Tap Tapered WP Single-Pat devices because a packaging sealing defect may compromise sterility during shelf life.

    Product
    Screw Tap Tapered WP Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Tapered Dental Drills Due to Sterility Concerns

    Nobel Biocare is recalling 260 units of tapered dental drills because a packaging sealing issue may compromise sterility during shelf life.

    Product
    Drill Tapered RP 4.3x11.5 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Implant Retrieval Instruments Recalled for Sterility Concerns

    Nobel Biocare is recalling 11 units of Implant Retrieval Instrument Tri Channel 6.0 22 mm because a packaging sealing issue may compromise sterility.

    Product
    Implant Retrieval Instrument Tri Channel 6.0 22 mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Dental Drills Recalled Due to Packaging Seal Defects

    Nobel Biocare is recalling 264 tapered dental drills because a packaging sealing defect may compromise sterility during shelf life.

    Product
    Drill Tapered WP 5x10 mm Single-Pat - Product Usage: Drills are intended to be used in the upper and lower jaw bone to prepare osteotomy prior to implant placement.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Abutment Screw Retrieval Instruments Recalled for Sterility Concerns

    Nobel Biocare is recalling 233 units of Abutment Screw Retrieval Instruments because a packaging defect may compromise sterility during shelf life.

    Product
    Abutment Screw Retrieval Instrument 3.0/NP - Product Usage: The instruments for abutment screw retrieval can be used if an abutment screw is broken and a part of the screw is left inside the implant screw channel. If the implant thread needs to be cleaned before placing a new scr
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Dense Bone Drills Recalled for Sterility Concerns

    Nobel Biocare is recalling 8 Dense Bone Drills because a packaging defect may compromise sterility during shelf life.

    Product
    Dense Bone Drill 6.0 6x16 mm Single-Pat- Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Tapered Dental Drills Due to Sterility Concerns

    Nobel Biocare is recalling 8 units of tapered dental drills because a packaging sealing issue may compromise sterility during shelf life.

    Product
    Drill Tapered NP 3.5x8 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Implant Retrieval Instruments Due to Sterility Concerns

    Nobel Biocare is recalling 423 implant retrieval instruments because a packaging defect may compromise sterility during shelf life.

    Product
    Implant Retrieval Instrument CC NP & Ext Hex WP 22 mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Implant Retrieval Instrument Recall Due to Sterility Concerns

    Nobel Biocare is recalling 31 units of Implant Retrieval Instrument CC 3.0 31 mm because a packaging sealing defect may compromise sterility.

    Product
    Implant Retrieval Instrument CC 3.0 31 mm- Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Implant Retrieval Instruments Recalled for Sterility Concerns

    Nobel Biocare is recalling 2,470 implant retrieval instruments because a packaging defect may compromise sterility. The devices are distributed worldwide.

    Product
    Implant Retrieval Instrument Hex & Tri-Ch NP/RP 22mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Implant Retrieval Kits Due to Sterility Concerns

    Nobel Biocare is recalling 87 units of Implant Retrieval Kits because a packaging sealing issue may compromise the sterility of the devices during their shelf-life.

    Product
    Implant Retrieval Kit - Product Usage: The kit itself does not have a specific intended Use, but included devices have specific intended use as shown above.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Abutment Screw Retrieval Instruments Recalled for Sterility Concerns

    Nobel Biocare is recalling 132 units of Abutment Screw Retrieval Instrument RP/WP/6.0 because a packaging sealing issue may compromise sterility during shelf life.

    Product
    Abutment Screw Retrieval Instrument RP/WP/6.0 - Product Usage: The instruments for abutment screw retrieval can be used if an abutment screw is broken and a part of the screw is left inside the implant screw channel. If the implant thread needs to be cleaned before placing a new
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Recalls Tapered Drills Due to Sterility Concerns

    Nobel Biocare is recalling 473 tapered dental drills because a packaging defect may compromise sterility during shelf life.

    Product
    Drill Tapered NP 3.5x11.5 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Dense Bone Drills Recalled for Sterility Concerns

    Nobel Biocare is recalling Dense Bone Drills because a packaging defect may compromise sterility during shelf life.

    Product
    Dense Bone Drill WP 5x13 mm Single-Pat - Product Usage: Drills are intended to be used in the upper and lower jaw bone to prepare osteotomy prior to implant placement.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Implant Retrieval Instruments Recalled for Sterility Concerns

    Nobel Biocare is recalling 197 implant retrieval instruments because a packaging defect may compromise sterility during shelf life.

    Product
    Implant Retrieval Instrument Hex & Tri-Chi NP/RP 31mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Dental Drills Recalled Due to Sterility Concerns

    Nobel Biocare is recalling 59 dental drills because a packaging defect may compromise sterility. The devices are distributed worldwide.

    Product
    Drill Tapered WP 5x11.5 mm Single-Pat - Product Usage: Drills are intended to be used in the upper and lower jaw bone to prepare osteotomy prior to implant placement.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2020-02-26

    Nobel Biocare Implant Retrieval Instruments Recalled for Sterility Concerns

    Nobel Biocare is recalling 526 Implant Retrieval Instruments because a packaging sealing issue may compromise sterility during shelf life.

    Product
    Implant Retrieval Instrument CC RP & Tri Ch Hex WP 22 mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connecti
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-26

    Nobel Biocare Tapered Drills Recalled Due to Sterility Concerns

    Nobel Biocare is recalling 154 units of Tapered RP 4.3x13 mm drills because a packaging defect may compromise sterility.

    Product
    Drill Tapered RP 4.3x13 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
    Category
    Distribution
    0 states