The Recall Desk

FDA (Devices) recall action 84274

Polymer Technology Systems, Inc. recalled 4 products on 2020-01-29

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2020-01-29
Critical
0
Units
177,792

Why these products were recalled

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • SevereFDA (Devices)··2020-01-29

    Polymer Technology Systems Recalls CardioChek Plus Analyzers Due to Battery Overheating Risk

    Polymer Technology Systems is recalling 44,448 CardioChek Plus Professional Analyzers because a battery terminal defect can cause reverse contact and potential overheating.

    Product
    Henry Schein CardioChek Plus Professional Analyzer, New Analyzer, boxed, with carrying case, Reference Number 4060 HS, 570-0411
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-01-29

    PTS Diagnostics CardioChek Plus Analyzer Recalled for Battery Overheating Risk

    Polymer Technology Systems is recalling 44,448 CardioChek Plus Professional Analyzers because a battery terminal defect can cause overheating if batteries are inserted incorrectly.

    Product
    PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-01-29

    PTS Diagnostics CardioChek Plus Analyzer Recalled for Battery Overheating Risk

    Polymer Technology Systems is recalling 44,448 refurbished CardioChek Plus Professional Analyzers because a battery terminal defect may cause overheating if batteries are inserted incorrectly.

    Product
    PTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Numbers 2702
    Category
    Distribution
    0 states