The Recall Desk

FDA (Devices) recall action 84183

Haemonetics Corporation recalled 22 products on 2020-02-05

The FDA (Devices) publishes a recall like this one as 22 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 22 pages.

Products
22
Announced
2020-02-05
Critical
0
Units
1,753,721

Why these products were recalled

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–22 of 22

  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver 5/5+ Bowl Kits Due to Leak Risk

    Haemonetics is recalling Cell Saver 5/5+ Bowl Kits because they may leak, trapping fluid and causing incomplete washing.

    Product
    Cell Saver¿5/5+ Bowl Kit (225 ml) Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls CS5/5+ Bowl Kits Due to Potential Leaks

    Haemonetics is recalling CS5/5+ Bowl Kits because potential leaks may trap fluid, causing device errors and incomplete washing.

    Product
    CS5/5+ Bowl Kit-125 mL Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Japan Disp Sets Due to Potential Bowl Cracks

    Haemonetics is recalling specific Japan Disp Sets for Cell Saver systems due to potential bowl cracks that may cause leaks and incomplete blood washing.

    Product
    JAPAN DISP SET, 261J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls CS Elite Processing Sets Due to Potential Bowl Cracks

    Haemonetics is recalling CS Elite Processing Sets (125 ml) because cracks in the bowl core may cause fluid trapping and incomplete blood washing.

    Product
    CS Elite Processing Set (125 ml) Autotransfusion Device - Product Usage:its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and p
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Bowl Sets Due to Leak Risk

    Haemonetics is recalling specific Cell Saver bowl sets because they may leak, potentially trapping fluid and causing incomplete washing.

    Product
    Japan DISP SET, 263J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls CS Elite Fastpack Autotransfusion Devices Due to Leak Risk

    Haemonetics is recalling CS Elite Fastpack autotransfusion devices because cracks in the bowl core may cause leaks and incomplete washing.

    Product
    CS Elite Fastpack, 125 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Bowl Sets Due to Leak Risk

    Haemonetics is recalling Cell Saver 5/5+ and Elite bowl sets because they may leak, trapping fluid and causing incomplete washing.

    Product
    High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Elite Sets Due to Potential Bowl Leaks

    Haemonetics is recalling Cell Saver Elite sets because potential leaks in the bowl core may cause fluid trapping and incomplete washing.

    Product
    JAPAN CELL SAVER ELITE SET - 125ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processe
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver 5/5+ Fastpacks Due to Potential Bowl Cracks

    Haemonetics is recalling Cell Saver 5/5+ Fastpacks because cracks in the bowl core may cause fluid trapping and incomplete washing.

    Product
    CS5/5+ Fastpacks, 125 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls CS5/5+ Fastpacks Due to Potential Bowl Cracks

    Haemonetics is recalling CS5/5+ Fastpacks because cracks in the bowl core may trap fluid, causing device errors and incomplete washing.

    Product
    CS5/5+ Fastpacks, 225 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Elite Sets Due to Potential Bowl Cracks

    Haemonetics is recalling Cell Saver Elite sets because cracks in the bowl core may trap fluid, causing incomplete washing of recovered blood.

    Product
    JAPAN CELL SAVER ELITE SET - 225ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processe
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver 5/5+ Fastpacks Due to Potential Leaks

    Haemonetics is recalling 1,340 units of Cell Saver 5/5+ Fastpacks because potential leaks in the inner core may cause fluid trapping and incomplete washing.

    Product
    CS5/5+ Fastpacks, 125 mL, 20¿ Res Autotransfusion Device- Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Fastpacks Due to Potential Bowl Cracks

    Haemonetics is recalling 8,641 Cell Saver 5/5+ Fastpacks because cracks in the bowl core can trap fluid, causing incomplete washing and device errors.

    Product
    CS5/5+ Fastpacks, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls CS Elite Fastpack Autotransfusion Devices Due to Leak Risk

    Haemonetics is recalling CS Elite Fastpack autotransfusion devices because cracks in the bowl core may cause leaks and incomplete blood washing.

    Product
    CS Elite Fastpack, 225 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, an
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Bowl Sets Due to Potential Leaks

    Haemonetics is recalling Cell Saver bowl sets because cracks may cause leaks and incomplete blood washing.

    Product
    Japan CS Elite FASTPACK DISP. SET - 125ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this p
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Bowl Sets Due to Leak Risk

    Haemonetics is recalling 739 Cell Saver bowl sets because cracks in the inner core can cause leaks and incomplete blood washing.

    Product
    JAPAN DISP SET, 261J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Japan DISP SETs Due to Potential Bowl Cracks

    Haemonetics is recalling 1,584 Japan DISP SETs because cracks in the bowl core may cause leaks and incomplete washing.

    Product
    Japan DISP SET, 263J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls CS Elite Processing Sets Due to Potential Bowl Cracks

    Haemonetics is recalling CS Elite Processing Sets because cracks in the bowl core may cause fluid trapping and incomplete washing.

    Product
    CS Elite Processing Set (225 ml) Autotransfusion Device - Product Usage:its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and p
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls CS Elite Fastpack Autotransfusion Devices Due to Leak Risk

    Haemonetics is recalling CS Elite Fastpack autotransfusion devices because cracks in the bowl core may cause leaks and incomplete blood washing.

    Product
    CS Elite Fastpack, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, an
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Elite Fastpacks Due to Potential Bowl Cracks

    Haemonetics is recalling Cell Saver Elite Fastpacks because cracks in the bowl core may cause fluid trapping and incomplete blood washing.

    Product
    JAPAN CS ELITE FASTPACK, 225ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed re
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver Bowl Sets Due to Leak Risk

    Haemonetics is recalling Cell Saver bowl sets because cracks in the inner core may cause leaks and incomplete blood washing.

    Product
    JAPAN BASIC HI SPEED CS5 SET,LN 263J - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and was
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2020-02-05

    Haemonetics Recalls Cell Saver 5/5+ and Elite Bowl Sets Due to Leak Risk

    Haemonetics is recalling Cell Saver 5/5+ and Elite bowl sets because potential leaks may trap fluid, causing incomplete washing of recovered blood.

    Product
    JAPAN LOW VOL(125ML) CS5 SET,LN 261J - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and was
    Category
    Distribution
    50 states