The Recall Desk

FDA (Devices) recall action 83990

Isopure Corp recalled 3 products on 2020-01-29

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2020-01-29
Critical
0
Units
24

Why these products were recalled

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2020-01-29

    Isopure Centrifugal Pump Recall Due to Magnet Encapsulation Defect

    Isopure Corp is recalling eight Finish Thompson Centrifugal Pumps due to a manufacturing defect that could expose magnets to solution, causing corrosion and iron contamination.

    Product
    Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)
    Category
    Distribution
    13 states
  • HighFDA (Devices)··2020-01-29

    Isopure Acidified Concentrate Distribution System Recalled for Magnet Defect

    Isopure Corp is recalling 15 Acidified Concentrate Distribution and Storage Systems due to a potential magnet encapsulation defect that could cause corrosion and iron exposure.

    Product
    Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3
    Category
    Distribution
    13 states
  • HighFDA (Devices)··2020-01-29

    Isopure Sodium Bicarbonate Mixing System Recalled for Magnet Encapsulation Defect

    Isopure Corp is recalling one IsoMix PRO system due to a potential magnet encapsulation defect that could cause corrosion and iron exposure.

    Product
    Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3
    Category
    Distribution
    13 states