FDA (Devices) recall action 83980

Respironics California, LLC recalled 4 products on 2020-02-19

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2020-02-19
Critical
0
Units
3,897

Why these products were recalled

Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchscreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.

Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2020-02-19

    Respironics V60 Ventilator Recalled for Frozen Touchscreen Failure

    Respironics is recalling 2,851 V60 Ventilators because the touchscreen may freeze, preventing clinicians from adjusting therapy settings. This could delay necessary changes, potentially leading to hypercarbia or hypoxemia.

    Product
    Respironics V60 Ventilator, Service # 850008, REF: 1053613, 1053614, 1053617, 1053614
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-19

    Respironics Ventilator Touchscreen Recall Due to Unresponsive Controls

    Respironics is recalling V60/V680 ventilator touchscreens that may freeze and fail to respond to touch commands, potentially delaying necessary therapy adjustments.

    Product
    Respironics RP-Touch Screen, V60/V680 Ventilator Serviced with Touchscreen RP Kit 453561511951 containing a touchscreen labeled as raw part number 1132318 Revision C
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-02-19

    Respironics V680 Ventilator Recalled for Unresponsive Touchscreen

    Respironics is recalling 190 V680 Ventilators because the touchscreen may freeze, preventing clinicians from adjusting therapy settings and potentially causing delayed treatment.

    Product
    Respironics V680 Ventilator, Service # 850011, (OUS Only)
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-02-19

    Respironics V60 Ventilator User Interface Kits Recalled for Touchscreen Failure

    Respironics is recalling V60 Ventilator User Interface kits because the touchscreen may freeze, preventing clinicians from adjusting therapy settings.

    Product
    Respironics RP-UI Assy, V60 Ventilators User Interface RP-Kits, REF/P/N: 453561528931, 453561528941, 453561536451, 453561536461