The Recall Desk

FDA (Devices) recall action 83610

Pacific Medical Group Inc. recalled 10 products on 2020-02-19

The FDA (Devices) publishes a recall like this one as 10 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 10 pages.

Products
10
Announced
2020-02-19
Critical
0
Units
47,774

Why these products were recalled

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–10 of 10

  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Toco M2734A Fetal Monitors Recalled for Unverified Components

    Pacific Medical Group is recalling 466 Toco M2734A fetal monitors serviced with unverified non-OEM components that may fail to detect fetal distress or cause burns.

    Product
    Toco M2734A, that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Loop-Style Tocotransducers Recalled for Unverified Service Components

    Pacific Medical Group is recalling Loop-Style Tocotransducers serviced with unverified non-OEM components, which may cause inaccurate fetal monitoring or patient burns.

    Product
    Loop-Style Tocotransducer (Nautilus) 2264LAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Toco Board, NFCM9310; Pacific Medical GE Corometrics Nautilus TOCO Bottom Case, NFCM9315; Pacific Medical GE Corometrics Nautilus TOCO Cable…
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Group Recalls Ultrasound Transducers Serviced with Non-OEM Parts

    Pacific Medical Group is recalling 6,481 Button-Style Ultrasound Transducers serviced with unverified non-OEM components. The defect may cause inaccurate fetal and maternal monitoring or patient burns.

    Product
    Button-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700HAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Ultrasound Board, NFCM9510; Pacific Medical GE Corometrics Nautilus Ultrasound Bottom Case, NFCM9515, Pacific Medical GE…
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Ultrasound Transducers Recalled for Non-OEM Component Use

    Pacific Medical Group is recalling 5,897 ultrasound transducers serviced with unverified non-OEM components, which may cause inaccurate monitoring or patient burns.

    Product
    Ultrasound Transducer (M2736A), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips M2736A Avalon Smart Ultrasound Bottom Case, NFPHA9515
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical External Toco Transducers Recalled for Unverified Components

    Pacific Medical Group is recalling 1,603 External Toco Transducers serviced with non-OEM parts that were not appropriately verified or validated.

    Product
    External Toco Transducer (M1355A), that may have been serviced with the following: Pacific Medical Philips M1355A Toco Bottom Case, NFPH9315; Pacific Medical Philips M1355A Toco Cable Assembly, NFPH9320; Pacific Medical Philips M1355A Toco Gauge, NFPH9340
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Toco MP Transducers Recalled for Non-OEM Components

    Pacific Medical Group Inc. is recalling Toco MP Transducers serviced with unverified non-OEM components. The defect may cause inaccurate fetal monitoring and patient burns.

    Product
    Toco MP Transducer (M2734B), that may have been serviced with the following: Pacific Medical Philips M2734B Avalon Smart TOCO Bottom Case, NFPHA9315-B
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Toco+ Transducers Recalled for Unverified Service Components

    Pacific Medical Group is recalling 2,170 Toco+ transducers serviced with unverified non-OEM components. The defect may cause inaccurate fetal/maternal monitoring and patient burns.

    Product
    Toco+ transducer with ECG/IUP capability (wired), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Group Recalls Ultrasound Transducers Serviced with Non-OEM Parts

    Pacific Medical Group is recalling Loop-Style Ultrasound Transducers serviced with unverified non-OEM components. The parts may cause inaccurate fetal monitoring and patient burns.

    Product
    Loop-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700LAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Ultrasound Board, NFCM9510; Pacific Medical GE Corometrics Nautilus Ultrasound Bottom Case, NFCM9515, Pacific Medical GE…
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Ultrasound Transducers Recalled for Unverified Service Components

    Pacific Medical Group is recalling 1,994 ultrasound transducers serviced with unverified non-OEM components, which may cause inaccurate fetal and maternal monitoring.

    Product
    Ultrasound Transducer (M1356A), that may have been serviced with the following: Pacific Medical Philips M1356A Ultrasound Board, NFPH9510; Pacific Medical Philips M1356A Ultrasound Bottom Case, NFPH9515; Pacific Medical Philips M1356A Ultrasound Cable Assembly, NFPH9520; Pacific…
    Category
    Distribution
    51 states
  • SevereFDA (Devices)··2020-02-19

    Pacific Medical Group Recalls Button-Style Tocotransducers Due to Unverified Components

    Pacific Medical Group is recalling 9,679 Button-Style Tocotransducers serviced with non-OEM components that were not appropriately verified, posing risks of inaccurate monitoring and burns.

    Product
    Button-Style Tocotransducer (Nautilus), 8-foot Cord 2264HAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Toco Board, NFCM9310; Pacific Medical GE Corometrics Nautilus TOCO Bottom Case, NFCM9315; Pacific Medical GE Corometrics Nautilus…
    Category
    Distribution
    51 states