The Recall Desk

FDA (Devices) recall action 82691

Boston Scientific Corporation recalled 9 products on 2019-05-29

The FDA (Devices) publishes a recall like this one as 9 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 9 pages.

Products
9
Announced
2019-05-29
Critical
0
Units
69

Why these products were recalled

Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–9 of 9

  • HighFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga XL 4007 Laser Systems for Cooling Defect

    Boston Scientific is recalling 27 Auriga XL 4007 laser systems worldwide due to a mechanical cooling issue that may cause the device to stop emitting laser pulses during surgery.

    Product
    AURIGA XL 4007 GENERAL SYSTEM UPN: M0068FS4007G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incis
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga 30 Laser Systems Due to Power Degradation

    Boston Scientific is recalling specific Auriga 30 laser systems because a mechanical cooling issue may cause faster-than-normal power degradation and low power errors.

    Product
    AURIGA 30 BRAZIL ZERO COST SYSTEM UPN: M0068S30BZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform inc
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga XL 4007 Laser Systems Due to Power Degradation

    Boston Scientific is recalling three Auriga XL 4007 laser systems because a mechanical cooling issue may cause faster-than-normal power degradation and low power errors.

    Product
    AURIGA XL 4007 GENERAL - DEMO SYSTEM UPN: M0068FS4007GD0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perfo
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga 30 Laser Systems Due to Power Degradation

    Boston Scientific is recalling 18 Auriga 30 laser systems worldwide because a mechanical cooling issue may cause faster-than-normal power degradation and low power errors.

    Product
    AURIGA 30 GENERAL SYSTEM UPN: M0068S30G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, exc
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga 30 Laser Systems Due to Power Degradation

    Boston Scientific is recalling eight Auriga 30 laser systems worldwide because a mechanical cooling issue may cause faster-than-normal power degradation and low power errors.

    Product
    AURIGA 30 BRAZIL SYSTEM UPN: M0068S30B0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, exci
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga 30 Laser Systems Due to Power Degradation

    Boston Scientific is recalling two Auriga 30 laser systems because a mechanical cooling issue may cause faster-than-normal power degradation and a low power error.

    Product
    AURIGA 30 GENERAL ZERO COST SYSTEM UPN: M0068S30GZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform i
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga XL 4007 Laser System for Cooling Defect

    Boston Scientific is recalling specific Auriga XL 4007 laser systems due to a mechanical cooling issue that may cause premature power degradation and low power errors.

    Product
    AURIGA XL 4007 BRAZIL SYSTEM Model: M0068FS4007B0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform inc
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga 30 Laser Systems Due to Power Degradation

    Boston Scientific is recalling specific Auriga 30 laser systems because a mechanical cooling issue may cause faster-than-normal power degradation and low power errors.

    Product
    AURIGA 30 GENERAL - DEMO SYSTEM UPN: M0068S30GD0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform inci
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2019-05-29

    Boston Scientific Recalls Auriga XL 4007 Laser Systems for Cooling Defect

    Boston Scientific is recalling six Auriga XL 4007 laser systems worldwide due to a mechanical cooling issue that may cause power degradation and stop laser emission.

    Product
    AURIGA XL 4007 GENERAL - ZERO COST SYSTE UPN: FS4007GZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perfo
    Category
    Distribution
    0 states