FDA (Devices) recall action 82178

O-Two Medical Technologies, Inc. recalled 3 products on 2019-04-10

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2019-04-10
Critical
1
Units
887

Why these products were recalled

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • CriticalFDA (Devices)··2019-04-10

    O-Two Medical e600 Ventilator Recalled for Fire Risk

    O-Two Medical is recalling 119 e600 Automatic Transport Ventilators because a manufacturing defect may cause the unit to catch fire.

    Product
    e600 Automatic Transport Ventilator, Model Number 01EVE600
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2019-04-10

    O-Two Medical e700 Ventilator Recalled for Fire Risk

    O-Two Medical Technologies is recalling 632 e700 Automatic Transport Ventilators because a manufacturing defect could cause the unit to catch fire.

    Product
    e700 Automatic Transport Ventilator, Model Number 01EVE700
    Category
    Distribution
    8 states
  • SevereFDA (Devices)··2019-04-10

    O-Two Medical e500 Ventilator Recalled for Fire Hazard

    O-Two Medical is recalling 136 e500 Automatic Transport Ventilators because a manufacturing defect may cause the unit to catch fire.

    Product
    e500 Automatic Transport Ventilator, Model Number 01EVE500
    Category
    Distribution
    8 states