Randox Laboratories Ltd. recalled 3 products on 2019-03-27

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2019-03-27
Critical
0
Units
359

Why these products were recalled

The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • ModerateFDA (Devices)··2019-03-27

    Randox Fructosamine Control 3 Calibrator Recall Due to Value Reassignment

    Randox Laboratories is recalling 133 kits of Fructosamine Control 3 because the assigned value was decreased by 25%. The product is used for in vitro calibration.

    Product
    Fructosamine Control 3, FR2996 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.
    Category
    Distribution
    4 states
  • ModerateFDA (Devices)··2019-03-27

    Randox Fructosamine Calibrator Recall Due to Value Reassignment

    Randox Laboratories is recalling 141 kits of Fructosamine Control 1 because the assigned value was decreased by 25%.

    Product
    Fructosamine Control 1, FR2994 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.
    Category
    Distribution
    4 states
  • ModerateFDA (Devices)··2019-03-27

    Randox Fructosamine Calibrator Recall Due to Value Reassignment

    Randox Laboratories is recalling 85 kits of Fructosamine Calibrator FR2993 because the assigned value was decreased by 25%.

    Product
    Fructosamine Calibrator, FR2993 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.
    Category
    Distribution
    4 states