The Recall Desk

FDA (Devices) recall action 81680

GE Healthcare, LLC recalled 3 products on 2018-12-19

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2018-12-19
Critical
0
Units
7

Why these products were recalled

When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • SevereFDA (Devices)··2018-12-19

    GE Healthcare Recalls Discovery NM/CT 670 DR Due to Electrical Hazard

    GE Healthcare is recalling four Discovery NM/CT 670 DR units because incorrect wiring of safety switches may prevent the system from shutting off power, creating an electrical hazard.

    Product
    Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and…
    Category
    Distribution
    5 states