FDA (Devices) recall action 81238

Cook Medical Incorporated recalled 39 products on 2018-11-07

The FDA (Devices) publishes a recall like this one as 39 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 39 pages.

Products
39
Announced
2018-11-07
Critical
0
Units
383,076

Why these products were recalled

The instruction for use detail, with respect to removal of the delivery system as part of Deployment of the stent is incorrect and the IFU is being corrected to amend the instruction for use detail.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–25 of 39

  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 30mm, Catalog Number: ZIB6-80-5.0-30 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent for Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instructions for use contain incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 12mm x 40mm, Catalog Number: ZIB6-125-12.0-40 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Instruction Error

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 60mm, Catalog Number: ZIB6-125-5.0-60 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • LowFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 14mm x 60mm, Catalog Number: ZIB6-80-14.0-60 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 518 Biliary Stent Due to Instruction Error

    Cook Medical is recalling 9,834 Zilver 518 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 518 Biliary Stent 4mm x 40mm, Catalog Number: ZIB5-125-4.0-40 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 20mm, Catalog Number: ZIB6-125-5.0-20 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • LowFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent for Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 80mm, Catalog Number: ZIB6-80-5.0-80 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 4mm x 80mm, Catalog Number: ZIB6-125-4.0-80 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • LowFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 60mm, Catalog Number: ZIB6-80-5.0-60 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instructions for use contain incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 40mm, Catalog Number: ZIB6-125-5.0-40 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Instruction Error

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 30mm, Catalog Number: ZIB6-125-5.0-30 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 12mm x 60mm, Catalog Number: ZIB6-125-12.0-60 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • LowFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 14mm x 40mm, Catalog Number: ZIB6-125-14.0-40 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instructions for use contain incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 12mm x 80mm, Catalog Number: ZIB6-125-12.0-80 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 518 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 518 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 518 Biliary Stent 5mm x 20mm, Catalog Number: ZIB5-125-5.0-20 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • LowFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 518 Biliary Stent for Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 518 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 518 Biliary Stent 5mm x 80mm, Catalog Number: ZIB5-80-5.0-80 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 14mm x 30mm, Catalog Number: ZIB6-125-14.0-30 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 518 Biliary Stents Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 518 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 518 Biliary Stent 4mm x 20mm, Catalog Number: ZIB5-125-4.0-20 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 4mm x 80mm, Catalog Number: ZIB6-80-4.0-80 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 518 Biliary Stent for Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 518 Biliary Stents because the instructions for use contain incorrect details regarding the removal of the delivery system.

    Product
    Zilver 518 Biliary Stent 5mm x 80mm, Catalog Number: ZIB5-125-5.0-80 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 518 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 518 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 518 Biliary Stent 4mm x 30mm, Catalog Number: ZIB5-125-4.0-30 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stents Due to Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 12mm x 40mm, Catalog Number: ZIB6-80-12.0-40 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent for Incorrect Instructions

    Cook Medical is recalling 9,834 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 5mm x 40mm, Catalog Number: ZIB6-80-5.0-40 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver Biliary Stent Due to Incorrect Removal Instructions

    Cook Medical is recalling 9,834 Zilver Biliary 518 Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver Biliary 518 Stent 5mm x 60mm, Catalog Number: ZIB5-125-5.0-60 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states
  • ModerateFDA (Devices)··2018-11-07

    Cook Medical Recalls Zilver 635 Biliary Stent Due to Incorrect Instructions

    Cook Medical is recalling 9,384 Zilver 635 Biliary Stents because the instruction for use contains incorrect details regarding the removal of the delivery system.

    Product
    Zilver 635 Biliary Stent 14mm x 80mm, Catalog Number: ZIB6-125-14.0-80 Product Usage: The Zilver 518 and 635 Biliary Stents are intended for palliation of malignant neoplasms in the biliary tree.
    Category
    Distribution
    49 states