The Recall Desk

FDA (Devices) recall action 81190

Tosoh Bioscience Inc recalled 2 products on 2018-12-19

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2018-12-19
Critical
0
Units
490

Why these products were recalled

Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occu…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • ModerateFDA (Devices)··2018-12-19

    Tosoh AIA-900 and AIA-2000 Analyzers Recalled for Incubator Slippage Risk

    Tosoh Bioscience is recalling AIA-900 and AIA-2000 analyzers due to a potential incubator slippage issue that may cause system errors and delays in patient test results.

    Product
    AIA-900Analyzer, Product code 022930, 022930R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-12-19

    Tosoh AIA-2000 and AIA-900 Analyzers Recalled for Incubator Slippage

    Tosoh Bioscience is recalling AIA-2000 and AIA-900 analyzers because a timing belt issue can cause incubator slippage, leading to system errors and delays in patient test results.

    Product
    AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step…
    Category
    Distribution
    0 states