The Recall Desk

FDA (Devices) recall action 80960

Siemens Medical Solutions USA, Inc recalled 3 products on 2018-11-07

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2018-11-07
Critical
0
Units
184

Why these products were recalled

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • SevereFDA (Devices)··2018-11-07

    Siemens Ysio Radiographic Systems Recalled for Potential Tube Arm Drop

    Siemens is recalling 142 Ysio radiographic systems because steel cables may rupture, causing the tube arm to drop unexpectedly during patient positioning.

    Product
    Ysio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.
    Category
    Distribution
    38 states
  • SevereFDA (Devices)··2018-11-07

    Siemens AXIOM Multix MT Patient Table Recalled for Cable Failure Risk

    Siemens is recalling 11 AXIOM Multix MT patient tables because steel cables may rupture, causing the arm to drop unexpectedly during patient positioning.

    Product
    AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Systems radiographic exposures of the entire body.
    Category
    Distribution
    38 states
  • SevereFDA (Devices)··2018-11-07

    Siemens Ysio MAX Radiographic Systems Recalled for Cable Failure Risk

    Siemens is recalling 31 Ysio MAX radiographic systems because steel cables in the lifting column may rupture, causing the arm to drop unexpectedly.

    Product
    Ysio MAX (model number 10762470) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used in radiographic and tomographic exposures of the whole body.
    Category
    Distribution
    38 states