Ortho Clinical Diagnostics Inc recalled 4 products on 2018-10-10

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2018-10-10
Critical
0
Units
79,855

Why these products were recalled

The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • ModerateFDA (Devices)··2018-10-10

    Ortho VITROS FSH Reagent Pack Recalled for Potential Biased Results

    Ortho Clinical Diagnostics is recalling VITROS FSH Reagent Packs because biotin supplements may cause biased test results at lower concentrations than indicated.

    Product
    VITROS Immunodiagnostic Products - FSH Reagent Pack (10758750000302) 1931922
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-10-10

    Ortho VITROS Folate Reagent Pack Recall for Biotin Interference

    Ortho Clinical Diagnostics is recalling VITROS Folate Reagent Packs because biotin supplements may cause biased test results at lower concentrations than stated.

    Product
    VITROS Immunodiagnostic Products - Folate Reagent Pack(10758750009237) 1513266
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-10-10

    Ortho VITROS TSH Reagent Packs Recalled for Potential Biotin Interference

    Ortho Clinical Diagnostics is recalling VITROS TSH Reagent Packs because biotin supplements may cause biased test results at lower concentrations than stated in the instructions.

    Product
    VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-10-10

    Ortho VITROS Prolactin Reagent Pack Recalled for Biotin Interference

    Ortho Clinical Diagnostics is recalling VITROS Prolactin Reagent Packs because biotin supplements may cause biased test results at lower concentrations than stated in the instructions.

    Product
    VITROS Immunodiagnostic Products - Prolactin Reagent Pack (10758750000111) 1849793
    Category
    Distribution
    0 states