The Recall Desk

FDA (Devices) recall action 80693

GE Medical Systems China Co., Ltd. recalled 3 products on 2018-09-05

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2018-09-05
Critical
0
Units
963

Why these products were recalled

On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2018-09-05

    GE Carestation 30 Anesthesia Systems Recalled for Gas Switch Defect

    GE Medical Systems is recalling 252 Carestation 30 anesthesia systems worldwide because a gas switch may unexpectedly change position, potentially causing hypoxia.

    Product
    GE Carestation 30 anesthesia systems
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-09-05

    GE 9100c NXT anesthesia systems recalled for gas switch defect

    GE Medical Systems is recalling 401 anesthesia systems worldwide because a gas switch may unexpectedly change position, potentially causing hypoxia.

    Product
    GE 9100c NXT anesthesia systems
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-09-05

    GE Carestation Anesthesia Systems Recalled for Gas Switch Defect

    GE Medical Systems is recalling 310 anesthesia systems worldwide because a gas switch may unexpectedly change position, potentially causing hypoxia.

    Product
    GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems
    Category
    Distribution
    0 states