The Recall Desk

FDA (Devices) recall action 80516

Siemens Healthcare Diagnostics, Inc. recalled 4 products on 2018-09-05

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2018-09-05
Critical
0
Units
14,942

Why these products were recalled

Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2018-09-05

    Siemens RAPIDPoint 500 Blood Gas Analyzer Recalled for Hydroxocobalamin Interference

    Siemens recalls RAPIDPoint 500 Blood Gas Analyzers because Hydroxocobalamin may interfere with test results, potentially leading to incorrect medical assessments.

    Product
    RAPIDPoint 500 Blood Gas Analyzer; 10492730, 1069855, 10696857, 10697306 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-09-05

    Siemens RAPIDLab 1265 Blood Gas Analyzer Recalled for Hydroxocobalamin Interference

    Siemens is recalling 2,357 RAPIDLab 1265 Blood Gas Analyzers because Hydroxocobalamin may interfere with test results, potentially leading to incorrect medical assessments.

    Product
    RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-09-05

    Siemens RAPIDPoint 405 Blood Gas Analyzer Recalled for Interference

    Siemens is recalling RAPIDPoint 405 Blood Gas Analyzers because Hydroxocobalamin may interfere with test results, potentially leading to incorrect medical assessments.

    Product
    RAPIDPoint 405 Blood Gas Analyzer; 10282093, 10310464, 10314817, 10317193, 10318999, 10320055, 10321238, 10322347, 10328278, 10328302, 10336784 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-09-05

    Siemens RAPIDLab Blood Gas Analyzer Recalled for Hydroxocobalamin Interference

    Siemens Healthcare Diagnostics is recalling 138 RAPIDLab 1245 Blood Gas Analyzers because Hydroxocobalamin may interfere with test results, potentially leading to incorrect medical assessments.

    Product
    RAPIDLab 1245 Blood Gas Analyzer; 10321844, 10337179, 10491393 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including…
    Category
    Distribution
    0 states