The Recall Desk

FDA (Devices) recall action 80434

Beckman Coulter Inc. recalled 7 products on 2018-11-07

The FDA (Devices) publishes a recall like this one as 7 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 7 pages.

Products
7
Announced
2018-11-07
Critical
0
Units
958

Why these products were recalled

Sorting-Drive software version 4.2 (released in October 2015) introduced a new option which included the Sorting-Drive Client that is used for manual printing of barcodes. When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.

Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports that this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.

Sorting-Drive software version 4.2 (released in October 2015) introduced a new option which included the Sorting-Drive Client that is used for manual printing of barcodes; version 4.2 was replaced by 4.2.1 (they both contain the same option for manual barcode printing). When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs.The firm reports this does not affect automated label creation or reprinting of individual labels an…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–7 of 7

  • LowFDA (Devices)··2018-11-07

    Beckman Coulter Sorting-Drive 4.2 Software Recalled for Barcode Errors

    Beckman Coulter is recalling 25 units of Sorting-Drive 4.2 software due to a defect that may generate incorrect barcode IDs when printing consecutively numbered tube labels.

    Product
    Software Package, Sorting-Drive 4.2 User Kit, Part No. B73488, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Automate 2500 Family of products. The system automates the process of sorting, decapping, and
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-11-07

    Beckman Coulter iPAW 4.2 Software Recalled for Barcode Labeling Errors

    Beckman Coulter is recalling the iPAW 4.2 User Kit software because a specific manual barcode printing function may generate incorrect IDs for sample tubes.

    Product
    Software Package, iPAW 4.2 User Kit, Part No. B73489, Manufactured by: Beckman Coulter Biomedical GmbH The iPAW 4.2 User Kit (PN B73489) includes the iPAW software application CD (version 4.2). This function is primarily used to copy an existing barcode ID to replace the labe
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-11-07

    Beckman Coulter Recalls Sorting-Drive Software Due to Barcode Errors

    Beckman Coulter is recalling Sorting-Drive software versions 4.2 and 4.2.1 because the manual barcode printing function may generate incorrect IDs for sample tubes.

    Product
    Software Package, Sorting-Drive 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95347, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Automate 2500 Family of products. . The Sorting-Drive 4.2.1 Use
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-11-07

    Beckman Coulter iPAW Software Recall for Barcode Label Errors

    Beckman Coulter is recalling iPAW 4.2.1 software due to a defect that may generate incorrect barcode IDs when printing consecutive tube labels.

    Product
    Software Package, iPAW 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95343, Manufactured by: Beckman Coulter Biomedical GmbH Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for m
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-11-07

    Beckman Coulter Sorting-Drive Software Recall for Barcode Printing Errors

    Beckman Coulter is recalling Sorting-Drive software versions 4.2 and 4.2.1 because a manual barcode printing feature may generate incorrect IDs for tube labels.

    Product
    Sorting-Drive Professional Part, Sorting-Drive Professional Kit, (Sorting Drive Software v4.2.1), Part No. A94144, Beckman Coulter, Inc. The Sorting-Drive Professional Kit (PN A94144) includes the Sorting-Drive 4.2.1 User Kit (PN B95347) which includes the Sorting-Drive 4.2
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-11-07

    Beckman Coulter Sorting-Drive Software Recalled for Barcode Label Errors

    Beckman Coulter is recalling Sorting-Drive software versions 4.2 and 4.2.1 because a manual barcode printing feature may generate incorrect tube label IDs.

    Product
    Sorting-Drive Product Part, Sorting-Drive Plus Kit (Sorting Drive Software v4.2.1), Part No. A89483, Beckman Coulter, Inc. The Sorting-Drive Plus Kit includes the Sorting-Drive 4.2.1 User Kit which includes the Sorting-Drive 4.2.1 Software DVD. . The Sorting-Drive software fa
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-11-07

    Beckman Coulter Recalls iPAW Software Due to Barcode Labeling Errors

    Beckman Coulter is recalling iPAW software versions 4.2 and 4.2.1 because a specific barcode printing feature may generate incorrect IDs for tube labels.

    Product
    Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client for specimen tracking and workload statistics or a physical workstation for manua
    Category
    Distribution
    0 states