FDA (Devices) recall action 80099

TriMed Inc. recalled 4 products on 2018-06-27

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2018-06-27
Critical
0
Units
2,755

Why these products were recalled

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2018-06-27

    TriMed Nitinol Staples Recalled Due to Missing Sterilization Audit Records

    TriMed Inc. is recalling 323 Nitinol staples because evidence of required quarterly sterilization dose audits could not be found, posing a potential infection risk.

    Product
    Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2018-06-27

    TriMed Recalls Nitinol Staples Due to Missing Sterilization Audit Records

    TriMed Inc. is recalling 532 Nitinol staples because quarterly sterilization dose audit records could not be found, posing a risk of patient infection.

    Product
    Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2018-06-27

    TriMed Nitinol Staple Recall Due to Missing Sterilization Audit Records

    TriMed Inc. is recalling 1,194 Nitinol staples because evidence of required quarterly sterilization dose audits could not be found, posing a potential infection risk.

    Product
    Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10; e) ES-15x12; f) ES-10x14; g) ES-10x15
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2018-06-27

    TriMed Recalls Nitinol Surgical Staples Due to Sterilization Record Gaps

    TriMed Inc. is recalling 706 units of Nitinol Surgical Staples because quarterly sterilization dose audits could not be verified, posing a potential infection risk.

    Product
    Nitinol Staple, Elastic Staple Small-Asymmetrical, Sterile R, Model #: a) ES-10x13x15; b) ES-10x16x19; c) ES-10x17x19; d) ES-12x13x15; e) ES-12x15x17; f) ES-12x16x19; g) ES-15x13x16
    Category
    Distribution
    14 states