Datascope Corporation recalled 28 products on 2018-05-16

The FDA (Devices) publishes a recall like this one as 28 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 28 pages.

Products
28
Announced
2018-05-16
Critical
0
Units
238

Why these products were recalled

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

26–28 of 28

  • HighFDA (Devices)··2018-05-16

    Datascope recalls custom tubing kits due to seal separation risk

    Datascope is recalling four custom tubing kits because a seal defect may allow air to enter the device, potentially reducing blood flow in ECMO circuits.

    Product
    ECC Small - Respiratory BEQ-TOP 5210, Custom Tubing Kit, Catalog No. 701055604
    Category
    Distribution
    13 states
  • SevereFDA (Devices)··2018-05-16

    Datascope recalls BEQ-TOP 5210 ECC Set custom tubing kits

    Datascope is recalling 18 custom tubing kits because a seal defect may cause balloon collapse, increasing resistance to flow in the venous line.

    Product
    BEQ-TOP 5210 ECC Set - Respiratory, Custom Tubing Kit, Catalog No. 701055604R01
    Category
    Distribution
    13 states