FDA (Devices) recall action 79428

Siemens Medical Solutions USA, Inc recalled 15 products on 2018-05-09

The FDA (Devices) publishes a recall like this one as 15 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 15 pages.

Products
15
Announced
2018-05-09
Critical
0
Units
19,515

Why these products were recalled

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–15 of 15

  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Emotion 16 CT Scanner Recalled for Dose Calculation Error

    Siemens Medical Solutions is recalling 1,301 SOMATOM Emotion 16 CT scanners because a software algorithm may incorrectly calculate radiation dose for head scans, potentially exposing patients to unnecessary radiation.

    Product
    SOMATOM Emotion 16 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Scope CT Scanner Recalled for Incorrect Radiation Dose Calculations

    Siemens Medical Solutions is recalling SOMATOM Scope CT scanners because the CARE Dose4D algorithm may incorrectly calculate tube current for head scans, potentially leading to unnecessary radiation exposure.

    Product
    SOMATOM Scope Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Emotion Duo CT Scanner Recalled for Radiation Dose Errors

    Siemens is recalling SOMATOM Emotion Duo CT scanners because a software algorithm may calculate incorrect radiation doses for head scans, potentially exposing patients to unnecessary radiation.

    Product
    SOMATOM Emotion Duo (2003) Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Perspective CT Scanner Recalled for Dose Calculation Error

    Siemens Medical Solutions is recalling 1,301 SOMATOM Perspective CT scanners because a software algorithm may incorrectly calculate radiation dose for head scans, potentially leading to unnecessary exposure.

    Product
    SOMATOM Perspective Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Sensation Open CT Scanner Recall for Dose Calculation Error

    Siemens is recalling 1,301 SOMATOM Sensation Open CT scanners because a software algorithm may incorrectly calculate radiation dose for head scans, potentially leading to unnecessary exposure.

    Product
    SOMATOM Sensation Open Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Sensation 64 CT Scanner Recalled for Incorrect Dose Calculations

    Siemens Medical Solutions is recalling 1,301 SOMATOM Sensation 64 CT scanners because the CARE Dose4D algorithm may calculate incorrect tube current for head scans, potentially leading to unnecessary radiation exposure.

    Product
    SOMATOM Sensation 64 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Emotion CT Scanner Recalled for Incorrect Dose Calculations

    Siemens is recalling 1,301 SOMATOM Emotion CT scanners because the CARE Dose4D algorithm may incorrectly calculate tube current for head scans, potentially causing unnecessary radiation exposure.

    Product
    SOMATOM Emotion (2003) Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Scope Power CT Scanner Recall for Incorrect Dose Calculations

    Siemens Medical Solutions is recalling SOMATOM Scope Power CT scanners due to potential incorrect radiation dose calculations for head scans.

    Product
    SOMATOM Scope Power Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Perspective 16 CT Scanner Recalled for Dose Calculation Error

    Siemens is recalling 1,301 SOMATOM Perspective 16 CT scanners because a software algorithm may incorrectly calculate radiation dose for head scans, potentially exposing patients to unnecessary radiation.

    Product
    SOMATOM Perspective 16 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Emotion 6 CT Scanner Recalled for Dose Calculation Error

    Siemens is recalling 1,301 SOMATOM Emotion 6 CT scanners because the CARE Dose4D algorithm may incorrectly calculate radiation dose for head scans, potentially leading to unnecessary exposure.

    Product
    SOMATOM Emotion 6 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Sensation 10 CT Scanner Recalled for Dose Calculation Error

    Siemens Medical Solutions is recalling 1,301 SOMATOM Sensation 10 CT scanners due to a software error that may cause unnecessary radiation exposure during head scans.

    Product
    SOMATOM Sensation 10 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Sensation 16 CT Scanner Recalled for Dose Calculation Errors

    Siemens is recalling 1,301 SOMATOM Sensation 16 CT scanners because the CARE Dose4D algorithm may incorrectly calculate tube current for head scans, potentially causing unnecessary radiation exposure.

    Product
    SOMATOM Sensation 16 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Spirit CT Scanner Recalled for Incorrect Dose Calculations

    Siemens Medical Solutions is recalling SOMATOM Spirit CT scanners because the CARE Dose4D algorithm may calculate incorrect tube current for head scans, potentially leading to unnecessary radiation exposure.

    Product
    SOMATOM Spirit Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Sensation 40 CT Scanner Recalled for Dose Calculation Error

    Siemens is recalling 1,301 SOMATOM Sensation 40 CT scanners because the CARE Dose4D algorithm may incorrectly calculate tube current for head scans, potentially causing unnecessary radiation exposure.

    Product
    SOMATOM Sensation 40 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-05-09

    Siemens SOMATOM Sensation Cardiac CT Scanner Recalled for Dose Calculation Error

    Siemens is recalling 1,301 SOMATOM Sensation Cardiac CT scanners because a software algorithm may incorrectly calculate radiation dose for head scans, potentially exposing patients to unnecessary radiation.

    Product
    SOMATOM Sensation Cardiac Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
    Category
    Distribution
    0 states