The Recall Desk

FDA (Devices) recall action 79408

Zimmer Biomet, Inc. recalled 11 products on 2018-04-04

The FDA (Devices) publishes a recall like this one as 11 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 11 pages.

Products
11
Announced
2018-04-04
Critical
0
Units
37,994

Why these products were recalled

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–11 of 11

  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System because a device performance review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System ACE TROCH NAIL 125 DEG. X 11MM ACE TROCH NAIL 130 DEG. X 11MM ACE TROCH NAIL 135 DEG. X 11MM ACE TROCH NAIL 125 DEG. X 13MM ACE TROCH NAIL 130 DEG. X 13MM ACE TROCH NAIL 135 DEG. X 13MM The Trochanteric Nail System is intended to treat…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System because a Device Performance Review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System STER TROC END CAP STER TROC END CAP 5MM STER TROC END CAP 10MM STER TROC END CAP 15MM STER TROC END CAP 20MM STER TROC END CAP 25MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System after a device performance review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System STER TROC LAG SCR 70 MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System after a device performance review failed to meet clinical survivorship criteria.

    Product
    ACE Trochanteric Nail System STER TROC NAIL 125 DEG X 9MM STER TROC NAIL 130 DEG X 9MM STER TROC NAIL 135 DEG X 9MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System after a Device Performance Review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System STER TROC NAIL 180X11X140 STER TROC NAIL 125 DEG X 11MM STER TROC NAIL 130 DEG X 11MM STER TROC NAIL 135 DEG X 11MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System because a Device Performance Review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System TROCH NAIL ST 13X180 125DG STRL TROCH NAIL ST 13X180 130DG STRL TROCH NAIL ST 13X180 135DG STRL The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Screws

    Zimmer Biomet is recalling ACE Trochanteric Nail System screws because a device performance review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System ATN AR SCREW 60MM STERILE ATN AR SCREW 65MM STERILE ATN AR SCREW 70MM STERILE ATN AR SCREW 75MM STERILE ATN AR SCREW 80MM STERILE ATN AR SCREW 85MM STERILE ATN AR SCREW 90MM STERILE ATN AR SCREW 95MM STERILE ATN AR SCREW 100MM STERILE…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System After Failed Performance Review

    Zimmer Biomet is recalling the ACE Trochanteric Nail System because a Device Performance Review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System STER L H TROC NAIL 320X9X125 STER L H TROC NAIL 340X9X125 STER L H TROC NAIL 360X9X125 STER L H TROC NAIL 380X9X125 STER L H TROC NAIL 400X9X125 STER L H TROC NAIL 420X9X125 STER L H TROC NAIL 440X9X125 STER L H TROC NAIL 460X9X125 STER L H…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System after a Device Performance Review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System STER TROC NAIL 180X13X140 STER TROC NAIL 125 DEG X 13MM STER TROC NAIL 130 DEG X 13MM STER TROC NAIL 135 DEG X 13MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System because a device performance review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System TROCH NAIL ST 11X180 125DG STRL TROCH NAIL ST 11X180 130DG STRL TROCH NAIL ST 11X180 135DG STRL The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls ACE Trochanteric Nail System Due to Performance Failure

    Zimmer Biomet is recalling the ACE Trochanteric Nail System after a Device Performance Review failed clinical analysis for survivorship.

    Product
    ACE Trochanteric Nail System ATN LAG SCREW 60MM STERILE ATN LAG SCREW 65MM STERILE ATN LAG SCREW 70MM STERILE ATN LAG SCREW 75MM STERILE ATN LAG SCREW 80MM STERILE ATN LAG SCREW 85MM STERILE ATN LAG SCREW 90MM STERILE ATN LAG SCREW 95MM STERILE ATN LAG SCREW 100MM…
    Category
    Distribution
    0 states