The Recall Desk

FDA (Devices) recall action 79385

Zimmer Biomet, Inc. recalled 2 products on 2018-03-21

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2018-03-21
Critical
0
Units
24

Why these products were recalled

Zimmer Biomet is conducting a medical device recall for two lots of Comprehensive Humeral stems. The lots were commingled during the manufacturing process and were etched with incorrect lot numbers. The size difference between the two stem sizes is 1.1 inches. It is visually recognizable by the user that the box label does not match the contents of the box.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2018-03-21

    Zimmer Biomet Recalls Humeral Stems Due to Incorrect Lot Number Etching

    Zimmer Biomet is recalling 12 Humeral Stems because two lots were commingled and etched with incorrect lot numbers, creating a risk of using the wrong implant size.

    Product
    Humeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
    Category
    Distribution
    5 states
  • ModerateFDA (Devices)··2018-03-21

    Zimmer Biomet Recalls Humeral Stems Due to Incorrect Lot Number Etching

    Zimmer Biomet is recalling 12 Humeral Stems because two lots were commingled and etched with incorrect lot numbers. The size difference is visually recognizable to the user.

    Product
    Humeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
    Category
    Distribution