The Recall Desk

FDA (Devices) recall action 79176

Tosoh Bioscience Inc recalled 2 products on 2018-03-07

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2018-03-07
Critical
0
Units
8,901

Why these products were recalled

Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • ModerateFDA (Devices)··2018-03-07

    Tosoh Bioscience Recalls ST AIA-PACK Progesterone III Assay Kits

    Tosoh Bioscience is recalling ST AIA-PACK Progesterone III Assay kits because cross-reactivity with DHEA-S supplements can cause falsely elevated test results.

    Product
    ST AIA-PACK¿ PROG III, Progesterone III Assay, Part Number: 025240
    Category
    Distribution
    39 states