FDA (Devices) recall action 79100

Steris Corporation recalled 29 products on 2018-03-07

The FDA (Devices) publishes a recall like this one as 29 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 29 pages.

Products
29
Announced
2018-03-07
Critical
0
Units
1,870

Why these products were recalled

On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 29

  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Tubing Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Tubing because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temperature Tubing 9"x100'. One roll per box; 10 boxes per case 6 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temperature Tubing 4"x100'. One roll per box; 10 boxes per case 10 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 5"x15" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris is recalling Vis-U-All High Temp Heat Seal Pouches because purple ink may migrate onto the pouch exterior or the medical devices inside after steam processing.

    Product
    Vis-U-All High Temp 3"x8" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temperature Tubing 14"x100'. One roll per box; 10 boxes per case 6 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Pouches Due to Ink Migration

    Steris is recalling Vis-U-All High Temp Self Seal Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and ink transfer to devices.

    Product
    Vis-U-All High Temp 13"x18" Self Seal Pouch 100 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 6"x10" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Pouches Due to Ink Migration Defects

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 7.5"x13" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to devices.

    Product
    Vis-U-All High Temp 16"x16" Self Seal Pouch 100 pouches per box; 6 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 18"x22" Self Seal Pouch 100 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 10"x15" Heat Seal Pouch 100 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 8"x16" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to devices.

    Product
    Vis-U-All High Temp 3.5"x22" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temp Heat Seal Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and ink transfer to devices.

    Product
    Vis-U-All High Temp 16"x16" Heat Seal Pouch 100 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 4.5"x11" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 3.5"x9" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Tubing Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Tubing because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temperature Tubing 6"x100'. One roll per box; 10 boxes per case 10 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 18"x22" Heat Seal Pouch 100 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Pouches Due to Ink Migration Defects

    Steris is recalling Vis-U-All High Temp Self Seal Pouches because purple ink may migrate onto the pouch exterior or the medical devices inside after steam processing.

    Product
    Vis-U-All High Temp 4"x11" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to enclosed devices.

    Product
    Vis-U-All High Temp 7.5"x13" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temp Heat Seal Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 8"x10" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temp Heat Seal Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 7.5"x13.5" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All High Temperature Sterilization Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to enclosed devices.

    Product
    Vis-U-All High Temp 5.5"x10" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temperature Sterilization Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 8"x16" Heat Seal Pouch 100 pouches per box; 9 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2018-03-07

    Steris Recalls Vis-U-All Heat Seal Pouches Due to Ink Migration

    Steris Corporation is recalling Vis-U-All High Temp Heat Seal Pouches because purple ink on the exterior label may migrate during steam processing, causing cosmetic defects and potential ink transfer to medical devices.

    Product
    Vis-U-All High Temp 4"x8" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed…
    Category
    Distribution
    0 states