Siemens Medical Solutions USA, Inc recalled 12 products on 2017-09-06

The FDA (Devices) publishes a recall like this one as 12 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 12 pages.

Products
12
Announced
2017-09-06
Critical
0
Units
11,832

Why these products were recalled

Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–12 of 12

  • ModerateFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen x-ray systems recalled for cooling leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen x-ray systems in the US because a cooling system leak could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US due to a cooling system leak that could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US because a cooling system leak could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US due to a cooling system leak that can cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US due to a cooling system leak that could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US due to a cooling system leak that could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US due to a cooling system leak that can cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US because a cooling system leak could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems due to a cooling system leak that could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US due to a cooling system leak that can cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2017-09-06

    Siemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak

    Siemens is recalling 986 AXIOM Artis zee and Q/Q.zen angiography systems in the US due to a cooling system leak that can cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2017-09-06

    Siemens Medical Solutions Recalls AXIOM Artis zee and Q Angiography Systems

    Siemens Medical Solutions is recalling 986 AXIOM Artis zee and Q angiography systems in the US due to a cooling system leak that could cause equipment failure.

    Product
    Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese…
    Category
    Distribution
    0 states