FDA (Devices) recall action 77761

Smiths Medical ASD Inc. recalled 2 products on 2018-01-10

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2018-01-10
Critical
0
Units
19,574

Why these products were recalled

Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2018-01-10

    Smiths Medical Medfusion Syringe Infusion Pump Recall for Clamp Mechanism Defect

    Smiths Medical ASD Inc. is recalling Medfusion Syringe Infusion Pumps because a loose clamp mechanism may cause the device to incorrectly alarm for invalid syringe sizes.

    Product
    Medfusion, Syringe Infusion Pump, Software Version V1.1, Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the administration of fluids requiring precisely controlled infusion rates including blood or blood…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2018-01-10

    Smiths Medical Medfusion Syringe Infusion Pump Recall for Clamp Loosening

    Smiths Medical is recalling Medfusion Syringe Infusion Pumps because a loose clamp mechanism may cause the device to incorrectly alarm for invalid syringe sizes.