FDA (Devices) recall action 76293

Siemens Healthcare Diagnostics, Inc. recalled 2 products on 2017-02-22

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2017-02-22
Critical
0
Units
13,147

Why these products were recalled

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • ModerateFDA (Devices)··2017-02-22

    Siemens Enzyme II Calibrator Recalled for ALTI Result Shifts

    Siemens Healthcare Diagnostics is recalling Enzyme II Calibrators because they may cause patient and QC results to shift by up to 10% on the Dimension clinical chemistry system.

    Product
    The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry system. For prescription use only.
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2017-02-22

    Siemens Enzyme 2 Calibrator Recalled for Positive Result Shifts

    Siemens Healthcare Diagnostics is recalling Enzyme 2 Calibrators because they may cause patient and QC results to shift up to 10% when used with the Dimension Vista System.

    Product
    The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
    Category
    Distribution
    0 states