The Recall Desk

FDA (Devices) recall action 76255

Resource Optimization & Innovation Llc recalled 135 products on 2017-04-05

The FDA (Devices) publishes a recall like this one as 135 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 135 pages.

Products
135
Announced
2017-04-05
Critical
0
Units
39,233

Why these products were recalled

Custom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

51–75 of 135

  • HighFDA (Devices)··2017-04-05

    FDA Recalls Sterile Custom Procedure Trays Due to Glove Splitting Risk

    Resource Optimization & Innovation LLC is recalling 660 sterile custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800326002, Sterile, GS0757B - Major Basin - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays Due to Splitting Light Glove Component

    Resource Optimization & Innovation Llc is recalling 12 custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800549002, Sterile, ET0638B - Head & Neck - Medical Center Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Resource Optimization Recalls 15 Sterile Custom Procedure Trays

    Resource Optimization & Innovation LLC is recalling 15 sterile custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800526003 Sterile, ET0610C - Head and Neck - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays with Defective Light Gloves

    Resource Optimization & Innovation LLC is recalling 100 custom procedure trays because the included light gloves can split, breaking the sterile field.

    Product
    Regard, Item Number: 800551002, Sterile, GS0640B - Lap Chole - Medical Center Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Robotic Hysterectomy Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation LLC is recalling 206 sterile robotic hysterectomy procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800567003, Sterile, GY0656C - Robotic Hysterectomy - Alamo
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Custom Procedure Trays Due to Light Glove Splitting

    Resource Optimization & Innovation LLC is recalling 150 Regard custom procedure trays because the included light gloves can split, breaking the sterile field.

    Product
    Regard, Item Number: 800558003, Sterile, OR0647C - Extremity - Childrens Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Pacemaker Procedure Trays Due to Light Glove Splitting

    Resource Optimization & Innovation LLC is recalling 26 sterile pacemaker procedure trays because the included light gloves can split, breaking the sterile field.

    Product
    Regard, Item Number: 880328002, Sterile, HT0759B - Pacemaker Pk - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    FDA Recalls 460 Sterile Lap Chole Procedure Kits Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 460 sterile lap chole procedure kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800528005 Sterile, GS0612E - Lap Chole PK - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Lap Chole Kits Due to Glove Splitting

    Resource Optimization & Innovation LLC is recalling 834 sterile lap chole kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800528002 Sterile, GS0612B - Lap Chole PK - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Spohn Laparoscopy Kits Due to Light Glove Splitting

    Resource Optimization & Innovation LLC is recalling 92 Spohn laparoscopy kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800417002, Sterile, GY0486B - Laparoscopy Pk - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays with Defective Light Gloves

    Resource Optimization & Innovation LLC is recalling 30 custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800562001, Sterile, NU0651A - Neuro - Childrens Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays with Defective Light Gloves

    Resource Optimization & Innovation LLC is recalling 165 custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800559003, Sterile, ET0648C - Head & Neck - Childrens Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays with Defective Light Gloves

    Resource Optimization & Innovation LLC is recalling 189 custom procedure trays because the included light gloves can split, breaking the sterile field.

    Product
    Regard, Item Number: 800559002, Sterile, ET0648B - Head & Neck - Childrens Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Procedure Trays Due to Glove Splitting Hazard

    Resource Optimization & Innovation LLC is recalling 40 kits of Regard sterile procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800535002 Sterile, LD0620B - C Section - New Braunfels
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Total Knee Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation LLC is recalling 70 sterile total knee procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800553002, Sterile, OR0642B - Total Knee - Medical Center Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Lap Chole Procedure Kits Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 2,016 sterile lap chole procedure kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800409002, Sterile, GS0478B - Lap Chole PK - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Spohn AV Access Kits Due to Splitting Light Gloves

    Resource Optimization & Innovation LLC is recalling 306 Spohn AV Access Kits because the included light gloves can split, breaking the sterile field.

    Product
    Regard, Item Number: 800425001, Sterile, CV0494A - AV Access Pk - Spohn
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2017-04-05

    FDA Recalls Sterile Custom Procedure Trays Due to Glove Splitting Risk

    Resource Optimization & Innovation LLC is recalling 42 sterile custom procedure trays because a light glove component can split, potentially breaking the sterile field.

    Product
    Regard, Item Number: 880323003, Sterile, LD0754C - C Section - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Laparotomy Procedure Trays Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 30 sterile laparotomy procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 880342004, Sterile, GS0776D - Laparotomy - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Abdominal PK Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation Llc is recalling 200 sterile abdominal procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800533002 Sterile, GS0606B - Abdominal PK - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Resource Optimization Recalls Sterile Robotic Laparoscopic Procedure Trays

    Resource Optimization & Innovation LLC is recalling 41 sterile robotic laparoscopic procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800599003, Sterile, GS0694C - Robotic Lap - Christus Highland
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Sterile C-Section Procedure Kits Due to Glove Splitting Risk

    Resource Optimization & Innovation LLC is recalling 690 sterile C-section kits because a light glove component may split, breaking the sterile field.

    Product
    Regard, Item Number: 800408002, Sterile, LD0477B - C-Section PK - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of VP Shunt Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation LLC is recalling 55 VP Shunt procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800566002, Sterile, NU0655B - VP Shunt - Childrens Santa Rosa
    Category
    Distribution
    2 states